iBox Anesthesia System

8

Figure 6: Connecting the

Breathing Device Inside the

Darkroom.

Figure 7: Filling the Vaporizer with

Isoflurane Liquid

from the warming plate (see red circles, Figure 5). Then, attach the single nosecone to the
breathing device/manifold and insert the black plug into the nosecone to prevent gas from
escaping into the darkroom when not in use.

Inside the iBox Darkroom

1. Connect the hose on the breathing device/manifold (J)

marked ‘To Inlet’ (Figure 6) to the fitting marked ‘Inlet’
mounted inside the iBox.

2. Connect the hose on the breathing device/manifold (J)

marked ‘To Exhaust’ to the fitting marked ‘Exhaust’
mounted inside the iBox.

3. If not already attached, place the desired nosecones on

the port(s) of the breathing device/manifold, making
sure the black plugs are secure in the nosecone.

Vaporizer

The vaporizer must only be used with isoflurane or sevoflurane as indicated in the shipping
paperwork. No other agent is acceptable and may be dangerous.

NOTE: Isoflurane or sevoflurane should only be used by licensed personnel. Wear an
appropriate facemask and follow safety procedures as recommended by OSHA and facility
safety officers. Pregnant women should not be exposed to isoflurane. For questions regarding
safety, contact UVP for assistance.

During use, frequently check that the isoflurane or sevoflurane level is between the minimum
and maximum marks on the vaporizer’s sight glass level indicator (Figure 7). Refill the
vaporizer before the liquid level reaches the minimum mark.

The vaporizer must never be modified, dismantled, calibrated or serviced by unauthorized
personnel. The vaporizer must be serviced at an approved Service Center. The unit should be
serviced for cleaning and calibration every one to two years depending on laboratory
requirements.

The vaporizer must be connected so that the flow of gas to the animal is as indicated by the
arrows on the device. The incorrect flow direction will result in wrong dosage.

Before use, ALL connections must be checked for leaks and functional tests must be
performed, as is the normal procedure for an anesthetic device.

WARNING: Keep the vaporizer upright at all times and do not carry the vaporizer by holding
the dial control.


To Fill the Vaporizer (Figure 7):

NOTE: The control dial must be in OFF position.
Press the white lock button to turn the dial. Always
keep the dial in the OFF position when not in use.
Also, the drain plug must be closed.

1. Remove the Filler Cap by turning counterclockwise.

2. Verify that the liquid agent is the correct type for the

system (isoflurane or sevoflurane).

3. Slowly pour the agent into the opening.

4. Observe the level through the sight glass on the

front of the vaporizer.

5. Once filled, replace the filler cap and tighten.


Drain Plug

Sight Glass Level Indicator

Filler Cap