Service procedures, Return procedure, Replacement parts and accessories – UVP GelDoc-Ite Imager User Manual

Page 11: Troubleshooting

Advertising
background image

GelDoc-It

e

Imager

11

Service Procedures

Return Procedure

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service prior to
returning any product.

Replacement Parts and Accessories

To order accessories or replacement parts for the GelDoc-It

e

Imager, contact UVP’s offices.

Part Description

Part Number


Fuses:

Fuse, 3.2A (for Darkroom)

56-0002-05 (Qty. 2 Required)

Fuse, 2A (for 115V Transilluminators)

56-0002-01 (Qty. 2 Required)

Fuse, 2A (for 230V Transilluminators)

56-0002-03 (Qty. 2 Required)

Emission Filters:

Filter, Ethidium Bromide, 50mm Square

38-0220-01

Filter, SYBR Green, 50mm Square

38-0219-01

Filter, SYBR Gold, 50mm Square

38-0221-01

Transillumination Accessories:

White Light Converter Plate, 21x26cm

38-0191-01

Visi-Blue Converter Plate, 21x26cm

38-0200-01

Gel Accessories:

Gel-Cutter

85-0002-01

Gel-Ruler

85-0003-01

Gel-Scooper

85-0006-01

Gel-Tray, small

85-0007-01

Gel-Sentry DNA Preparation Plate

97-0076-01

Fluorescent Standard Step Tablet

33-0014-02

Protective Equipment:

Spectacles, UV Blocking (UVC-303)

98-0002-01

Goggles, UV Blocking (UVC-503)

98-0002-02

Faceshield, UV Blocking (UVC-803)

98-0002-04

Thermal Printer and Paper:

Thermal Printer

89-0069-06

High Gloss, High Contrast Thermal Paper (5 Rolls per Box)

89-0174-01

Matte Finish, High Contrast Thermal Paper (4 Rolls per Box)

89-0038-01

Troubleshooting

No Power to the Darkroom or Transilluminator

1. Recheck the main power cord connection to the GelDoc-It

e

darkroom as well as the jumper cable

between the darkroom and transilluminator.

2. Check the fuses located at the back of the unit next to the power port. A flat-head screwdriver will

be required.

Turn the fuseholder cap counterclockwise and the fuse holder will pop out. Inspect the thin wire
within the glass fuse to see if there is a break in the wire. If so, replace the fuse(s). If fuses are
blowing repeatedly, contact UVP Technical Support for additional troubleshooting.

Advertising