Setup Instructions

Components

When unpacking the iBox Spectra, the following items will be included:

1. iBox Spectra darkroom
2. Manual BioLite MultiSpectral Light Source
3. Warming plate
4. 455-495nm excitation filter and 515LP emission filter
5. UVP flash drive
6. Power and USB cables
7. Keyboard and mouse
8. Supporting documentation


When unpacking and setting up the system, two people are required to move the darkroom.

Place the darkroom on a flat surface which can provide adequate support for up to 100 pounds.

WARNING: Do not attempt to perform any setup procedures while the system is plugged in or powered on
unless otherwise instructed.

CAUTION: Do not install the system in areas with high moisture, dust or high temperatures. Keep the
equipment away from motors or any other large magnetic equipment apparatus. This system is designed for
indoor use only.

Connecting the Power Cables

1. Plug the main power cable into the back of the

darkroom and the other end into a surge-
protected power outlet.

Note: It is recommended to leave the power
switch on the back of the system in the ON
position except when the system will not be
used for an extended period of time (one day or
longer).

Note: Do not position the system so that it is difficult to access the power cable and operate the
main power switch at the back of the unit.

2. If installing optional epi handheld lamps, place the lamps in the brackets located at the top of the

darkroom. Plug the handheld lamps into the outlets provided inside the darkroom. Place the power
switches located on the outside of the handheld lamps in the ON position.

Installing the Warming Plate

1. Place the warming plate inside the darkroom on top of the Roll-Out Warming Plate Tray.

2. Plug the jumper cord into the outlet located in the upper left back panel of the darkroom.

3. Plug the other end into the warming plate.

NOTE: When the warming plate is plugged into the outlet in the darkroom and the darkroom is turned
on, the warming plate will receive a constant supply of power.