Replacing the tube in uv/air circulator, Technical assistance, Warranty – UVP Ultraviolet PCR User Manual

Page 14

Advertising
background image

Ultraviolet PCR Cabinets and Workstations

Page 14

Replacing the Tube in UV/Air Circulator

1. Locate and remove the two thumb knobs on

the circulator cap on the outer back panel of
the system.


2. Pull the circulator cap away from the system.

3. The tube, attached to the cap, has two prongs

on each end that fit into a socket. Carefully
hold the tube on the ends, rotate the tube and
remove from the cap. Insert a new tube into the
socket. Twist the tube 90° until the tube locks
into place.

4. Replace the place cap back into position and reattach the two thumb knobs.

Technical Assistance

UVP offers technical support for all of its products. If you have any questions about the product’s
use, operation or repair, please contact our offices at the locations below.

Note: A Returned Goods Authorization (RGA) number must be obtained from UVP Customer
Service before returning any product.

You may contact UVP’s Customer Service based on your location:

If you are in North America, South
America, East Asia or Australia:

If you are in Europe, Africa, the
Middle East or Western Asia:

Call (800) 452-6788 or (909) 946-3197, and
ask for Customer Service during regular
business days, between 7:00 a.m. and 5:00
p.m. PST.

Call +44(0) 1223-420022, and ask for
Customer Service
during regular
business days, between 8:30 a.m. and
5:30 p.m.

E-Mail your message to: [email protected]

E-Mail

your

message

to:

[email protected]

Fax Customer Service, and send it to:

(909) 946-3597.

Fax Customer Service, and send it to:

+44(0)1223-420561.

Write to: UVP, LLC 2066 W. 11

th

Street,

Upland, CA 91786 USA

Write to: Ultra-Violet Products Ltd.,

Unit 1, Trinity Hall, Nuffield Road,
Cambridge CB4 1TG UK

Warranty

UVP, LLC warrants its products to be free of detects in materials and workmanship for a period of
two (2) years from the date of purchase. The foregoing warranty of UVP shall be of no force and
effect if buyer has modified or damaged the product. Tubes and filters are warranted for 90 days.

NOTE: “Crazing” (cracking and/or yellowing) of the Makrolon

®

panels may occur within the

warranty period. This is regarded as normal wear and tear and is not covered by the warranty.

All warranties or merchantability and fitness for any purpose and all other warranties, expressed or
implied, except those expressly set forth herein, are deemed waived and excluded.

UVP’s duty under the warranty is limited to replacement and/or repair of the defective part at the
option of UVP, LLC. UVP shall not be liable for any expenses or damages incurred by the
purchaser except as expressly set forth herein, and in no event shall UVP be liable for any special,
incidental or consequential damages of any kind. This warranty does not supersede any statutory
rights that may be available in certain countries.

Makrolon is a registered trademark of Bayer AG.

Circulator

cap with

UV tube

Advertising