Maintenance, Check and maintenance, Exchange of pre-treatment cartridge – Yamato Scientific WG221S Water Purifiers User Manual

Page 41: Exchange of ion exchange resin cartridge

Advertising
background image

-39-

Maintenance

Check and maintenance

Time for periodical check and maintenance

(Check periodically the unit to use in stable

condition.)

Item to check and maintain

Period

Note

Exchange of pre-treatment cartridge

Each 6 or 12
months

Durability:
About 5000 in case of using water service
in Tokyo (with quality of 200×10

-4

S/m)

You need earlier exchange of the cartridge
when you use raw water with worse quality.

Exchange of ion-exchange resin cartridge

When the
indicator (1)
shows

Durability:
About 650 of raw water with quality of
200×10

-4

S/m

Cleaning the still

3 months

Cleaning the water supply hose filter

6 months

You need earlier cleaning when you use raw
water with worse quality.

Hose exchange

2 years

Check the joint once a month.

Distilled water tank

3 months

Exchange of Pre-Treatment Cartridge

Refer to Page 15 “Connect securely the pre-treatment cartridge” in the “Safety Precaution”.
Dispose of the old pre-treatment cartridge as nonflammable.
If you use the unit without changing the pre-treatment cartridge, the span of life of the ion-exchange resin
cartridge is made short.

Exchange of Ion Exchange Resin Cartridge

Because the keeping of the cartridge for long time makes the water quality and durability worse, prepare
the spare cartridge for the time to exchange. The standard of keeping in about four months.
Refer to Page 14 “Connect securely ion-exchange resin cartridge” in the “Safety Precaution”.
Dispose of the exchanged cartridge as nonflammable.
If you use the unit without changing the cartridge, stone attaches increasingly to the inside of the boiler
and heaters, which can decrease collection of distilled water and damage heaters.