Collecting and Reviewing Data Chapter 5

Start Data Collection

After setting the data collector options and selecting a route, you are now ready to
start data collection.

Follow these steps to start collecting data.

1.

Once you have selected a list, you begin collecting data from the designated
point on the route.

2.

Connect the transducer and press Enter to start a measurement.

3.

If necessary, press Enter again to collect a measurement depending on the
Auto Store setting in the Instrument Configuration screen.

See

Instrument Configuration Parameters on page 59

.

You may notice that a Bias Check is taken if the parameter is set to Auto.

IMPORTANT

The Dynamix 2500 data collector supplies power to ICP
accelerometers through the A and B connectors. The ICP power `turns
on’ prior to collecting the first measurement point and remains on
until the following occurs:

You reach the end of the route.

The data collector advances to a measurement point that does not

require ICP power.

You exit the route menu.

Do not connect the channel to a BNC panel or non ICP device until the
Dynamix 2500 data collector advances to a non ICP measurement
point. Connecting the channel while collecting a measurement point
that requires ICP power may cause undesirable results, damage to the
equipment, or unexpected machinery trips.