External connections, Connector a and connector b inputs, Power/usb/ trigger laser – Rockwell Automation 1441-DYN25-Z Dynamix 2500 Data Collector User Manual

Page 19

Advertising
background image

Rockwell Automation Publication 1441-UM001B-EN-P - September 2012

19

The Dynamix 2500 Data Collector Chapter 1

External Connections

The external hardware connections for communication and data collection are
located on the top panel of the Dynamix 2500 data collector.

Figure 2 - External Hardware Connections

Connector A and Connector B Inputs

The data collector has two signal inputs (LEMO connectors), connector A and
B. Headphone access is supported on Connectors A and B.

The measurement Input signal range is ±25 V maximum. Input over voltage
protection is AC ± 50 V peak, DC ± 50 V. The inputs are protected ±50 V AC
or DC sustained against high-voltage transients, but trigger range over-voltage
input levels must be avoided.

Voltage can be DC or AC coupled, while the third option, Accel (ICP) 24 V DC
@ 2.4mA, is available for direct connection of integrated circuit piezoelectric
transducers. These settings are specified in the Emonitor software.

See

Setting Up Measurements on page 65

for more information.

A

POWER/USB/

TRIGGER

LASER

B

32144-M

1

5

6

7

4

3

2

1

5

6

7

4

3

2

1

2

3
4

5

6

7

Table 5 - Dynamix 2500 Data Collector Pin Assignments

Pin

Connector A

Connector B

POWER/USB/TRIG

1

Audio Output

Audio Output

USBV

2

Channel X

Channel Y

USB+

3

Channel Z

N/C

USB-

4

Strobe Out

Strobe Out

DIGITAL GND

5

ANALOG GND

ANALOG GND

EXT-DC-IN

6

Channel Y

N/C

±

25 V EXT-TRIG-IN

7

Channel R

Channel Z

+5V DC TACHO SUPPLY

TIP

In some circumstances you may find a reference to channel numbers
instead of X, Y, Z or R. Channel numbers can be cross referenced to
X= Ch(1), Y = Ch(2), Z= Ch(3), R = Ch(4).