Status indicators – Rockwell Automation 1441-DYN25-Z Dynamix 2500 Data Collector User Manual

Page 22

Advertising
background image

22

Rockwell Automation Publication 1441-UM001B-EN-P - September 2012

Chapter 1 The Dynamix 2500 Data Collector

Status Indicators

The Dynamix 2500 data collector has four status indicators located in the

upper-right side of the instrument nameplate.

Figure 4 - Status Indicators

Table 7 - Status Indicators

Status
Indicator

Operation

State

Definition

Dual State
Status

Red

Charging Battery

Flashing

When first connected to the power adapter, the
Dynamix 2500 Data Collector tests the condition of
the battery pack. Within 30 seconds, the status
indicators should go to a solid state.

If the status indicator continues to flash, the
instrument has diagnosed a fault in the battery pack
and it will not charge the battery. The anomaly may
be that the cell temperature is too high or there is a
fault with the battery pack.

Solid

The battery pack is OK. The power adapter is
connected to the instrument and charging. The
battery pack is maintained by a trickle charge from
the power adapter for as long as it remains attached
to the instrument.

Data Collection

Solid

An alarm condition has been set.
Input signal is out of range.
ICP check has failed and there is an ICP fault

condition.

The power adapter is connected to the

instrument.

Blue

Communication

Solid

Communicating to Emonitor through the USB

connection.

Amber

Data Collection

Solid

Acquisition sub-system is settling.

Green

Data Collection

Solid

Input signal is stable and data is ready to be stored.

32146-M

Dual Status: Red and Blue

Green

Amber