This figure shows the setup for a typical set of measurement definitions with
multiple frequency bands.

Figure 16 - Magnitude with Multiple Frequency Bands Measurement Definitions

(1)

The filter specifications are shown in the table below:

Magnitude and Phase Measurements at Orders

the Emonitor software and the Dynamix support collecting the magnitude and
phase at orders. This topic describes the steps to set up measurement definitions
to collect magnitude and phase with the Dynamix. This table describes items that
you may need to know about collecting magnitude and phase at orders.

(1) The filter specifications used in the tables are not currently defined in the Emonitor software database. To

define a new filter specification, choose Setup > Band Filters and click New.

Table 26 - Filter specifications for magnitude with multiple frequencies

Name

Low Cutoff

High Cutoff

Data collector filter type

0.8-1.2 Order

0.8 Orders

1.2 Orders

None

1.2-3.5 Order

1.2 Orders

3.5 Orders

None

3.5-8.5 Order

3.5 Orders

8.5 Orders

None

8.5-50% Fmax

8.5 Orders

50% Max Freq

None

50%-100% Fmax

50% Max Freq

100% Max Freq

None

0.8-100% Fmax

0.8 Orders

100% Max Freq

None

Item

Description

Collecting phase

You must select Also collect phase data in the collection

specification to collect phase.

The data collector can collect phase only at the first through eighth

orders.

Defining order filters

You must select Analog when you define orders filters. The Dynamix
2500 data collector automatically uses its internal order filters if you
are collecting phase.

If you are collecting only magnitude values and you do not select this
option, the data collector collects a spectrum and the Emonitor
software calculates the magnitudes after unloading.

Collecting magnitude and
phase at orders

When collecting magnitude and phase at orders, the data collector
always collects a spectrum measurement, and then calculates the
magnitude and phase values from the spectrum.