The following table shows the collection specifications (in the same order as the
measurement definition table above).

The following table shows the transducer specification in the collection
specifications.

Time Waveform Measurement Definitions

The Emonitor software and the Dynamic 2500 data collector support time
waveform measurement definitions. This topic lists the unique characteristics of
the Dynamix that you may need to know when setting up and collecting time
waveform data.

The Dynamix 2500 data collector cannot collect phase with time waveform
measurement definitions. You can choose to have the instrument collect a time
waveform only if a magnitude measurement at the same location is in alarm.

The Dynamic 2500 data collector supports integrated time waveform
measurements (single integration). This means you can collect a time waveform
with velocity units using an accelerometer or displacement units using a velocity
transducer.

The length of the time waveform depends on the maximum frequency you select
in the collection specification. You can calculate the number of time samples from
the number of spectral lines (Collection Specification dialog box) using this
formula.

You can calculate the total time required to collect a time waveform measurement
from the Number of lines value and the Maximum frequency value in the
Collection Specification dialog box (Setup > Collection Specification).

Table 32 - Collection specifications for spectrum measurement definitions

Transducer

Window

Signal Detection

Fmax

Lines

Phase

Order Norm?

Averages

Accelerometer

Hanning

Peak

2 kHz

400

No

No

4 linear

Accelerometer

Hanning

Peak

10 orders

400

No

Yes

4 linear

Accelerometer

Hanning

Peak

2 kHz

400

No

No

4 time synch.

Table 33 - Transducer specification in spectrum collection specifications

Name

Base Unit

Input Type

Units

Calibration Value

DC Offset

Accelerometer

Acceleration

ICP Accel

g’s

100

0

TIP

Set the Collect and store condition in the Edit Storage Specification
dialog box to On Magnitude Alarm for the storage specification for the
time waveform measurement definition.

time samples

spectral lines 2.56

=