FRF Frequency Response Function Extension
Module

The Frequency Response Function (FRF) extension module is an optional
module for the Dynamix 2500 data collector. You install the extension module
with the FRF Frequency Response Secure Digital (SD) storage card.

See

Install an Extension Module on page 8

for installation instructions.

The Dynamix 2500 data collector FRF extension module lets you determine the
natural frequencies of a machine or structure in a more sophisticated manner
than the Bump Test module. Rather than simply detecting the frequencies where
resonances occur, it captures information about the frequency response of the
structure being tested.

You can use the FRF data in conjunction with a modal analysis software

package to model the way the structure reacts to forces applied to it.

You can use the FRF module to capture information about how a machine

moves when it is running. You can use this information with the Operation
Deflection Shape (ODS) analysis software.

The FRF module includes the following features:

Simple selection of measurement type from a list of predefined options, for

example, mobility, stiffness, and apparent mass.

FRF trace colored green or red to illustrate high or low coherence.
Optional automatic parameter setting for input ranges, number of lines,

and windowing.

Compatibility with ODS analysis software.

Topic

Page

Newton’s Law and Apparent Mass

16

Setting Up an FRF Measurement

16

FRF Module Setup

18

Collect FRF Data

22

Field Calibration

25

Saving, Reviewing, and Deleting FRF Measurements

27

IMPORTANT

The FRF extension module is for Offroute/unscheduled measurements
only. The Emonitor software does not import this type of data.

Advertising