Operation, Maintenance and Troubleshooting Chapter 4

Reset Communications

If Ethernet port communications fail, the i-Sense has an assigned IP address that
may be incompatible with your current LAN. Reset the i-Sense IP address to the
default value by simultaneously pressing the RESET and Heartbeat

buttons for

10 seconds. This will also reset the security user name and password to their
default values. The Ethernet connection will restart after 5 seconds. Then follow
the

Ethernet Communications Setup on page 7

to assign a new IP address and

change username and password from default values.

Hard Reset

If the i-Sense fails to respond, press and hold the RESET buttons for 4 seconds to
perform a hard reset. This will not erase stored event data, but will cause a power
up event to be recorded.

Planned Shutdown

The i-Sense will normally record an interruption event whenever mains power is
disconnected. It is often preferable to exclude maintenance shutdowns from your
site’s power data record. To shut down the i-Sense without recording an
interruption event, push and hold the Heartbeat

button for 10 seconds, until

the LED blinks rapidly. The i-Sense will automatically restart the next time
power is cycled, or when the button is held for another 10 seconds.

SHOCK HAZARD: Do not remove the right cover while main voltage is applied.
The middle cover should never be removed.there are no user-serviceable parts
inside.

Battery Replacement

WARNING: To reduce risk of explosion or fire, replace only with same battery
type: Rechargeable NiMH, size AA, 1.2V, 2000-to-3000mAh. Obey polarity
markings (+ / -)

The i-Sense uses rechargeable batteries to allow continued operation during
power interruptions. The batteries will last for several years in normal operation,
but will need to be replaced at some point. Battery health is shown on the i-Sense
Management Console. The battery cover is located on the bottom surface of the
i-Sense. Battery replacement may be safely performed while main power is
applied. To replace or check batteries:

Loosen two Phillips-head screws and remove the battery cover to access
the battery holder.

Advertising