background image

9

Chapter 3

Operating Instructions

This chapter describes operator controls and indicators on the Bird 4421 RF
Power Meter. For remote operation using a GPIB or RS-232 controller, refer to
the instructions in Chapter 4 or Chapter 5 respectively.

Push Button Functions

Figure 7 Push Buttons

Push Button

Description

FWD, RFL

Press to measure forward or reflected RF power. FWD or RFL
indicator and current unit of measure turn on.

SWR

Press to measure standing wave ratio. SWR indicator turns on. Value
displayed will be between 1.0 and 199.9

MIN, MAX

Used after pressing FWD, RFL, SWR, or dBm. Displays the minimum
(maximum) measured value of the previous function as long as MIN
(MAX) is held down.

dBm

Used after pressing FWD or RFL. dBm indicator turns on. Power is
displayed in dBm units.



Note:

Used after pressing SWR. Return loss is displayed.

LIGHT

Press to turn on or turn off the display’s backlight. If left on, the light
automatically shuts off after 30 minutes.

AUTO

Press to automatically set the scale. AUTO turns on.

UP, DOWN

Press to select the next higher (lower) scale. If the scale is too high for
the power sensor, an error will be displayed.



Note:

Used while AUTO indicator is on. Stops automatic

scaling. AUTO indicator turns off.

ON/OFF

Press to turn the power meter on or off. This switch will not work if the
master power switch on the rear panel is OFF.



Note:

The display flashes for about 30 seconds when the

unit is first turned on.

LISTEN TALK LOCAL LOCKOUT

FWD

AUTO

LO BAT
REMOTE

MODEL 4421 RF POWER METER

AUTO

RANGE

UP

DOWN

SWR

RFL

FWD

MIN

MAX

dBm

LIGHT

ON / OFF

MW nW
KW µW
W mW
dBm

FWD
RFL
SWR