IEEE-488 GPIB Interface

15

Device CLear

(DCL)

Function

Resets the status of all devices to an initialized state.

Remarks

Does not change the current interface mode.

The 4421 returns to the factory default condition listed in Figure 5-6.

Table 5-6 IEEE-488 Default Conditions

Selective Device Clear

(SDC)

Function

Resets the status of a selected device to an initialized state.

Remarks

Only the device addressed will be cleared.

The 4421 returns to the factory default condition.

Group Execute Trigger

(GET)

Function

Initiates a measurement for all devices set to trigger on GET.

Remarks

The 4421 must be already set to trigger on GET.

Used to synchronize measurements of multiple instruments.

Default Condition

Related Command

Forward Carrier Wave

FC

Auto Range ON

RYY

Two Terminators (CR LF)

YT

Prefixes YES

PY

Trigger One Shot on Talk Address

T1

All SRQ’s OFF

M00

Send EOI with last byte of message

KY