Table 6-17 measurement ranges – Bird Technologies 4421 User Manual

Page 42

Advertising
background image

Bird 4421 RF Power Meter

30

MiNimum Value

(MN)

MaXimum Value

(MX)

Function

Selects minimum or maximum measurement mode.

Remarks

Another measurement must be selected before selecting min or max.
Returns the minimum (or maximum) value of the previous measurement type.

Range

(Rxx)

Function

Selects a measurement range listed in Figure 6-17.

Remarks

If the selected range is outside the range of the connected power sensor, the

command is ignored.

Table 6-17 Measurement Ranges

Terminators

(Yx)

Function

Selects the characters that follow the end of a data string. Set x to:

z

“T” for two terminators; a carriage return(CR) and a line feed(LF).

z

“O” for one terminator; a carriage return(CR).

z

“N” for no terminator.

Remarks

Many controllers use the terminator sequence to recognize the end of an input

string. Using incorrect terminators can lock the bus.

Prefixes

(Px)

Function

Turns the prefix mode on or off. Set x to:

z

“Y” to enable prefixes.

z

“N” to disable prefixes.

Command

Power Range

RYY

Turn auto range on

RNN

Turn auto range off. Keep present range

R17

18.0 – 199.9 MW

R16

1.80 – 19.99 MW

R15

0.180 – 1.999 MW

R14

18.0 – 199.9 kW

R13

1.80 – 19.99 kW

R12

0.180 – 1.999 kW

R11

18.0 – 199.9 W

R10

1.80 – 19.99 W

R09

0.180 – 1.999 W

R08

18.0 – 199.9 mW

R07

1.80 – 19.99 mW

R06

0.180 – 1.999 mW

R05

18.0 – 199.9 µW

R04

1.80 – 19.99 µW

R03

0.180 – 1.999 µW

R02

18.0 – 199.9 nW

R01

1.80 – 19.99 nW

R00

0.180 – 1.999 nW