5000-xt – Bird Technologies 5019D-Manual User Manual

Page 18

Advertising
background image

6

If calibration fails, “Cal Fail” will be displayed.

a.

Press a key to return to normal operation,

b.

Ensure that the WPS is properly connected, and the
RF is off.

c.

Rezero. Go back to Step 3.

5000-XT

1.

Make sure the sensor has reached a stable operating
temperature.

2.

Make sure no RF power is applied to the sensor.

3.

Press and hold “0” for two seconds to begin zero cali-
bration.

WARNING

Do not interrupt the calibration.

Note:

“Zero/Cal” will be displayed and calibration

will begin.
Note:

Zeroing the sensor takes 60 seconds. The

bar graph will display calibration progress.
Note:

When complete, “PASS” should be displayed.

4.

Press Enter to return to normal operation when zeroing is
complete.

Note:

If “FAIL” is displayed, make sure no RF power

is applied to the sensor and perform the procedure
again.