NIRS Vision Tutorial – Operation

Open Project “xdsstart”

1.

Click on the + sign next to the green cuvette icon.

Add Product to the Library

1.

Drag the product “AcDiSol” from the products folder into the library

folder.

2.

In the Library Product dialog box confirm the product ID and click on

[OK].

The product “AcDiSol” now appears under the library name next to a gray
(empty) cuvette. This indicates that “AcDiSol” has been chosen for sample
selection but has not yet completed the selection process.

Start Sample Selection

1.

Click on “AcDiSol”

2.

Click on Select  Apply Math

3.

In the Edit Sample Selection Parameters dialog box, enter the

following information:

Selected Method “Mahalanobis Distance in Principal Component

Space”

Math Treatment “2nd derivative”

Click on [OK]

4.

In the Edit Math Properties dialog box in the Principal Components

ID Parameters tab set the following wavelength regions:

Region 1

Minimum:

416

Maximum:

1080

Region 2

Minimum:

1120

Maximum:

2484

Cum Var.

95 %

Click on [OK]

5.

In the Sample Selection Options dialog box, set the following:

Find Outliers

enabled

Outlier Threshold

0.95

Threshold Type

Probability Level

Find Redundant

enabled

By number of samples

enabled

Training Set

75 %

Acceptance Set

25 %

Click on [OK]