Click on [ok, In the edit validation tab – Metrohm Vision – Tutorial User Manual

Page 33

Advertising
background image



5 Quantitative Model Development

26



NIRS Vision Tutorial – Operation

8.

Click on [OK].

9.

In the Regression Treatment Properties dialog box, in the

Wavelength Regions tab, enter the following settings:

Region 1

Minimum: 750

Maximum: 1050

Region 2

Minimum:1150

Maximum: 1350

Region 3

Minimum: 1500

Maximum: 1830

Region 4

Minimum: 1990

Maximum: 2450

10.

In the Edit Validation tab

Method

Cross Validation

Remove only one sample at a time

enabled

Click on [OK].

Vision performs calculations according to parameters defined in preceding
steps.

Evaluation of Method Performance

When the regression is complete, the screen displays 4 boxes in the Data View.
With statistical information relating to the Cross-validation results:

Figure 7

Result Partial Least Square Regression