Click on select ( apply math, Click on [ok – Metrohm Vision – Tutorial User Manual

Page 26

Advertising
background image

1.6 Multilinear Regression



NIRS Vision Tutorial – Operation



19

Perform Sample Selection

1.

Click on Select  Apply Math.

2.

In the Edit Sample Selection Parameters dialog box enter the

following settings:

Selection Method Mahalanobis Distance in Principal Component

Space

Math Treatment 2nd derivative

Leave all other settings unchanged

3.

Click on [OK].

4.

In the Edit Math Properties dialog box in the Principal Components

ID Parameters tab set the following wavelength regions:

Region 1

Minimum: 416
Maximum: 1080

Region 2

Minimum: 1120
Maximum: 2484

Cum Var. 95 %

Click on [OK]

5.

In the Edit Math Properties dialog box in the 2nd derivative tab set

the following values:

Segment: 10

Gap:

0

6.

In the Sample Selection Options dialog box apply or confirm the

following settings:

Find Outliers

enabled

Outlier Threshold

0.95

Threshold Type

Probability Level

Find Redundant

enabled

By number of samples

enabled

Training Set

75 %

Acceptance Set

25 %

Click on [OK]

Vision now performs calculations according to the parameters defined in the
previous steps.

Advertising