5.4 10-port valve

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

872 Extension Module – Liquid handling

5.4

10-port valve

5.4.1

Connecting the 10-port valve

Connecting the 10-port valve

NOTE

Connect all capillaries with 6.2744.010 long PEEK pressure screws:

1 Connection to the transfer tubing

Connect the capillary with a long pressure screw to the central
connector of the 10-port valve.

Fix the other end of the capillary to the coupling on the transfer
tubing.

2 Connecting transfer tubing and Dosino

Fix the capillary to the coupling of the transfer tubing.

Connect the other end of the capillary to the Dosino (see manual
for the Dosino)
.

3 Other capillary connections

The other 10 connectors can be freely used for:

Connection to the injection valve in the extension module.

Connection to the injection valve in the IC instrument.

Connection to the mixing vessel.

Connection to different solutions.

Connection to an ion exchanger.

etc.