5.6 Injection valve

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

872 Extension Module – Liquid handling

5.6

Injection valve

5.6.1

Connecting the injection valve

Connecting the injection valve

NOTE

Connect the capillaries and the sample loop with 6.2744.010 pressure
screws.

1 Connecting the sample path

Connect the sample inlet capillary to the connector 2.

Connect the sample outlet capillary to the connector 1.

2 Connecting the transfer solution path

Connect the transfer solution aspirating capillary to the connector
5.

Connect the transfer solution outlet capillary to the connector 4.

1

2

3

4

5

6

1

2

3

4

5

6

7

7

7

7

Figure 22

Injection valve - Measuring volume

1

Injection valve

2

Sample loop
For accurate measuring of small volumes.

3

Transfer solution outlet capillary
to the 10-port valve.

4

Transfer solution aspirating capillary
to the container with transfer solution.

5

Sample outlet capillary
to the waste container.

6

Sample inlet capillary
from the peristaltic pump

7

PEEK pressure screw 6.2744.010