7.7 Injection valve

■■■■■■■■■■■■■■■■■■■■■■

96

■■■■■■■■

930 Compact IC Flex ChS/PP (2.930.1300)

Safety shutdown

Function

Automatic shutdown upon reaching the pressure limit values

Maximum pres-
sure limit

Adjustable from 0.1 - 50 MPa (1 - 500 bar)

The pump is automatically shut down at the first piston stroke
above the maximum limit value

Minimum pres-
sure limit

Adjustable from 0 - 49 MPa (0 - 490 bar)

The shutdown mechanism is inactive at 0 MPa

The shutdown mechanism becomes active 2 minutes after system
start

The pump is automatically shut down after 3 piston strokes below
the minimum pressure limit

7.7

Injection valve

Actuator switch-
ing time

typ. 100 ms

Maximum operat-
ing pressure

35 MPa (350 bar)

Material

PEEK

7.8

Suppressor

Resistance to sol-
vents

No restriction

Switching time

typ. 100 ms

7.9

Peristaltic pump

Type

2-channel peristaltic pump

Direction of rota-
tion

Clockwise/counterclockwise

Rotational speed

0 - 42 rpm in 7 levels of 6 rpm each

Pumping charac-
teristics

0.3 mL/min at 18 rpm; with standard pump tubing (6.1826.420)

Pump tubing
material

Recommended: PharMed® (Ismaprene)

Advertising