7 servicing the inline filter, Servicing the inline filter – Metrohm 930 Compact IC Flex ChS/PP User Manual

Page 81

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

930 Compact IC Flex ChS/PP (2.930.1300)

■■■■■■■■

73

Connecting the input for the high-pressure pump

1

Remove the stopper from the coupling. Tighten the coupling to
the pressure screw located on the pump head input capillary (1).

Reconnect the pump head output capillary to the pump head out-
put (2).

5.7

Servicing the inline filter

NOTE

You can find a video sequence for this task in the Multimedia Guide IC
Maintenance
or on the Internet at

http://www.metrohm.com/com/

Support/Video/ICGeneral.html?q=14

.

Maintenance interval

The filter has to be replaced at least every 3 months; the filter has to be
replaced more frequently if you are working with high backpressure.

Accessories

For this task you need the following accessories:

Adjustable wrench (6.2621.000) from the accessory kit: Vario/Flex
Basic (6.5000.000)

Tweezers

A new filter from the packaging (6.2821.130)

Advertising