4 operation via rs232 interface, 1 general rules, 1 call up of objects – Metrohm 756 KF Coulometer User Manual

Page 61

Advertising
background image

4.1 General rules

756/831 KF Coulometer, Instructions of Use

53

4 Operation via RS232 Interface

4.1 General rules

The KF Coulometer has an extensive remote control facility that allows full control of

the KF Coulometer via the RS 232 interface, i.e. the KF Coulometer can receive data

from an external controller or send data to an external controller. CR and LF are used

as terminators for the data transfer. The KF Coulometer sends 2xCR and LF as ter-

mination of a data block, to differentiate between a data line which has CR and LF as

terminators. The controller terminates its commands with CR and LF. If more than

one command per line is sent by the controller, “;” is used as a separator between

the individual commands.

The data are grouped logically and easy to understand. Thus e.g., for the selection of

the dialog language, the following must be sent

&Config.Aux.Language "english"

whereby it is sufficient to only transmit the boldface characters, thus:

&C.A.L "english"

The quantities of the commands above are:

Config configuration

data

Aux

auxiliaries, various data

Language

setting the dialog language

The data are hierarchically structured (tree form). The quantities that occur in this

tree are called objects in the following. The dialog language is an object which can

be called up with the

&Config.Aux.Language

command.

If one is in the desired location in the tree, the value of the object can be queried.

&Config.Aux.Language $Q Q means Query

The query command $Q initiates the issuing of the value on the instrument and the

value emission is triggered. Entries which start with $, trigger something. They are

thus called triggers.

Values of objects can not only be queried, they can also be modified. Values are al-

ways entered in quotes, for example:

&Config.Aux.Language "english"

Advertising