Metrohm 756 KF Coulometer User Manual

Page 86

Advertising
background image

4.2 Remote control commands

756/831 KF Coulometer, Instructions for Use

78

4.2.2.27. Mode.CFmla

Mode.CFmla.1.Value

0...

±

999 999

Mode.CFmla.2.Value

etc., up to .19

Calculation constants specific to a method. Stored in the method memory of the

Coulometer. Operands specific to the sample (4.2.2.57 and 4.2.2.59) and values of

common variables (4.2.2.55) on the other hand are not stored with the methods.

4.2.2.28. UserMeth.FreeMem read

only

Memory space, available for user methods or silo lines. $Q sends the number of free

bytes, e.g.

"4928".

4.2.2.29. UserMeth.Recall $G

UserMeth.Recall.Name

up to 8 ASCII characters

UserMeth.Store $G

UserMeth.Store.Name

up to 8 ASCII characters

UserMeth.Delete $G

UserMeth.Delete.Name

up to 8 ASCII characters

UserMeth.DelAll $G

Management of the internal method memory: Load, store and delete methods. An

action is performed if "$G" is sent to the corresponding node just after entering the

name.

Do not use blank characters before and after method name!

.DelAll:

Deletes all methods in the user memory.

4.2.2.30. UserMeth.List.1.Name read

only

UserMeth.List.1.Mode read

only

UserMeth.List.1.Bytes read

only

UserMeth.List.1.Checksum read

only

for each method

List of the methods in the user method memory with the following characteristics:

.Name:

Name of the method

.Mode: Mode

.Bytes:

Number of bytes of the user memory used by the method

.Checksum: Checksum of the method, see 4.2.2.65.

4.2.2.31. Config.Monitoring.Reagent.Status ON,

OFF

Config.Monitoring.Reagent.Determ 1...99...999, OFF

Config.Monitoring.Reagent.DCounter

0...999

Config.Monitoring.Reagent.MaxTime 1...7...9999, OFF

Config.Monitoring.Reagent.TCounter

0...9999

Config.Monitoring.Reagent.ReagCap 1...1000...9999, OFF

Config.Monitoring.Reagent.RCounter

0...9999

Config.Monitoring.Reagent.ClearCount $G

Config.Monitoring.Reagent.Drift 0...99,

OFF

Advertising