6.1 Reference conditions

■■■■■■■■■■■■■■■■■■■■■■

104

■■■■■■■■

850 Professional IC –Anion – MCS – Gradient

6 Technical specifications

6.1

Reference conditions

The technical data listed in this Chapter refers to the following reference
conditions:

Ambient tempera-
ture

+25 °C (± 3 °C)

Instrument status

> 40 minutes in operation (equilibrated)

6.2

Instrument

IC system

Metal-free IC system

Compact system with modular design

Up to two complete chromatographic systems in one housing

Material

Painted polyurethane hard foam without CFCs, fire class V0

Operating pres-
sure range

0…50 MPa (500 bar) high pressure pump

0…35 MPa (350 bar) standard-PEEK system

Intelligent com-
ponents

iPump, iDetector, iColumn, MagIC Net

6.3

Leak sensor

Type

Electronic, no calibration necessary

6.4

Ambient conditions

Operation

Ambient tem-
perature

+5…+45 °C

Humidity

20…80 % relative humidity

Storage

Ambient tem-
perature

–20…+70 °C

Transport

Ambient tem-
perature

–40…+70 °C