3 procedure for checking instruments, 1 basics, 2 ph meters and titrators – Metrohm 767 Calibrated Reference User Manual

Page 10: Procedure for checking instruments, Basics, Ph meters and titrators, 3 procedure for checking instru- ments

Advertising
background image

3 Procedure for checking instruments

6

767 Calibrated Reference Instructions for Use

3 Procedure for checking instru-

ments

3.1 Basics

The 767 Calibrated Reference is connected instead of the sensors, if

possible by means of the original sensor cable. If this is not possible

(e.g. for electrodes without a plug-in head) a list of suitable cables can

be found in the appendix, see chap. 5.2.
Each instrument can be checked with the Calibrated Reference within

the normal operating program and therefore also with the worked-out

methods. This has the advantage that methods and selected function

runs can be tested at the same time.
On the other hand the Instructions for Use of most instruments contain

a so-called diagnosis instructions, selective instructions for checking

the functioning of the instrument if malfunctioning is suspected. This ty-

pe of check has the advantage that practically no (or only a very basic)

knowledge of operating the instrument is required. In addition, checking

the instrument by using the diagnosis instructions is usually significantly

faster.
This operating instructions for Calibrated Reference 767 are instructions

for checking the Metrohm instrument within the normal operating pro-

gram. As the very large range of instruments means that for individual

instruments many different names and operating structures may have

been used, these instructions should therefore be interpreted logically.
It is not absolutely necessary to firmly screw down the electrode cap at

sockets (4), (5), (6); plugging it in is quite adequate.

3.2 pH Meters and Titrators

Place the Calibrated Reference on the bench near the sensor. Ensure

that light is not hindered from reaching the solar cell (no shadows from

cables or accessories). If necessary switch on the room lighting.
On the instrument to be tested the slope must be set to 1, pH

as

to 7 and

the measuring temperature to 25 °C for measuring the pH.
Please note:
• On the basis of the pH calibration the pH is determined from the

measured potential value. The pH checked here is therefore chiefly

relevant as a functionality check.

• If the endpoint is evaluated from a curve in a titration, the absolute

measured potential or pH value is not relevant.

• With KF titrators this check should be evaluated as a functionality

test.

Advertising