3 Procedure for checking instruments

767 Calibrated Reference Instructions for Use

11

3. note following parameters, then set

(example 679) :
temperature (see above)

50°C

cond. range

20 µS/cm

paper feed

20 cm/h

4. press start

see G value for

Rancimat in certifi-
cate for 767.0010
(approx. 66 µS

1)

)

allow all channels

to write out 2 - 3 x
(the zero line is
shown in all chan-

nels)

5.

wait until the printer is print-
ing out a channel which has
not been checked.
Replug cable (see Fig. 8,
p. 11
), so that 14.3 kΩ ≅
approx. 69 µS is obtained

(see G value (5))

G value (5)
(ca. 69 µS

1)

)

allow all channels
to write out 2 - 3 x.
In the checked
channel the line will

be offset by the
amount of the al-
teration in conduc-
tance → check by

measuring with
ruler

6. if necessary repeat steps 1 - 5 for all

channels

1) Please consider the small number of decimal places in the display!








Fig. 7: Cable arrangement for 1st G value

(15.3 k

S

→approx. 66

:

S)

Fig. 8: Cable arrangement for 2nd G value

(14.3 k

S

→ approx. 69

:

S)

1

2

3

4

5

6

Rancimat

cable 6.2150.020

cable 6.2150.010

767

1

2

3

4

5

6

Rancimat

cable 6.2150.020

cable 6.2150.010

767