3 Procedure for checking instruments

767 Calibrated Reference Instructions for Use

7

3.2.1 U/mV,

pH

carry out on instrument or sen-

sor:

carry out on Calibrated

Reference:

compare display

with:

remarks

1. screw off cable at sensor (for plug-in

head electrodes, otherwise use
corresponding accessory cable, see
chap. 5.2

close cover

place sensor in

storage tube

2.

connect sensor cable to

socket (5)

3. measure mV

mV value (5)

4.

open cover

mV value (5)

compare with

permitted toler-
ance; note value

5.

connect sensor cable to
socket (4)

permitted

variation

from value noted

under step 4:
±0.1 mV (short-

time larger devia-
tions are normal)

6.

connect sensor cable to
socket (6)

mV value (6)

observe polarity;
(switch measuring
range if required);

compare with
permitted tolerance

7. measure pH

close cover

pH value (6)

set U

as

to pH 7 if

necessary

8.

open cover

pH value (6)

compare with

permitted tolerance

End of check

Steps 9...12 are of secondary importance. In general it is sufficient to carry out this check once per year.
For instruments with earthed circuits (e.g. all Titrinos and early series of 692/713) or for instruments without an earth socket (604,
704, 744) these steps are not relevant. Further Information to steps 9...12 can be found in chapter 2.5.

9. measure mV

connect sensor cable to
socket (5)

note display as
under step 4

10.

additionally connect cable

6.2150.020 (from accesso-
ries in case) to socket (6)

11. insert banana plug of cable in step

10 in earth socket of tested instru-
ment. Banana plug of shielding re-
mains open.

do not touch sockets (4),

(5), (6) during the meas-
urement

observe display

while connecting
the cable

permitted variation:
±0.1 mV display as
under step 5

(short-time larger
deviations are
normal)

12. remove cable from step 11

remove cable from

socket (6)

If the variation of the measured values is too large then first exchange

the original sensor cable against the reference cable in the accessories.

When the check is finished recalibrate the electrodes.