Metrohm 888 Titrando User Manual
Page 42
Advertising
6.2 SET titration
■■■■■■■■■■■■■■■■■■■■■■
34
■■■■■■■■
888 Titrando
Problem
Cause
Remedy
■
Select the user-defined titration rate and
decrease the control range.
■
Select the user-defined titration rate and
increase the maximum rate.
■
Select the user-defined titration rate and
increase the minimum rate.
The results are
spread widely.
The minimum dosing rate
is too high.
Select user-defined titration rate and decrease
the minimum rate (see manual/help of the
software used).
The electrode responds too
slowly.
Replace the electrode.