6.2 SET titration

■■■■■■■■■■■■■■■■■■■■■■

34

■■■■■■■■

888 Titrando

Problem

Cause

Remedy

Select the user-defined titration rate and
decrease the control range.

Select the user-defined titration rate and
increase the maximum rate.

Select the user-defined titration rate and
increase the minimum rate.

The results are
spread widely.

The minimum dosing rate
is too high.

Select user-defined titration rate and decrease
the minimum rate (see manual/help of the
software used).

The electrode responds too
slowly.

Replace the electrode.