8 Technical specifications

888 Titrando

■■■■■■■■

39

8 Technical specifications

8.1

Measuring interface

The 888 Titrando has one galvanically isolated measuring interface.

The measuring cycle is 100 ms for all measuring modes.

8.1.1

Potentiometry

One high-ohm measuring input (Ind.) for pH, metal or ion-selective elec-
trodes and one measuring input (Ref.) for separate reference electrodes.

Input resistance

> 1 · 10

12

Ω

Offset current

< 1 · 10

–12

A (under reference conditions)

Measuring mode
pH

Measuring
range

–13 - +20 pH

Resolution

0.001 pH

Measuring
accuracy

±0.003 pH
(±1 digit, without sensor error, under reference conditions)

Measuring mode
U

Measuring
range

–1,200 - +1,200 mV

Resolution

0.1 mV

Measuring
accuracy

±0.2 mV
(±1 digit, without sensor error, under reference conditions)

8.1.2

Temperature

A measuring input (Temp.) for temperature sensors of the Pt1000 or NTC
type with automatic temperature compensation.

R (25 °C) and B value can be configured for NTC sensors.

Measuring range

Pt1000

–150 - +250 °C

NTC

–5 - +250 °C
(R (25 °C) = 30,000

Ω and B (25/50) = 4,100 K)

Resolution

Pt1000

0.1 °C