5 Troubleshooting

856 Conductivity Module

■■■■■■■■

25

5 Troubleshooting

5.1

General

Problem

Cause

Remedy

Unstable measuring
signal - value near
zero.

The plug is plugged in
incorrectly.

The plug must be plugged in according to the
manual (dot to line).

Measured values dif-
fer strongly from
one another.

The cuvette is missing.

Screw the cuvette on to the 5-ring conductivity
measuring cell. Measure only with cuvette!

The measurement
signal is influenced
by the movement of
the 5-ring conductiv-
ity measuring cell
and/or its position in
the measuring ves-
sel.

The cuvette is missing.

Screw the cuvette on to the 5-ring conductivity
measuring cell. Measure only with cuvette!