7.1 Measuring interface

■■■■■■■■■■■■■■■■■■■■■■

30

■■■■■■■■

856 Conductivity Module

7 Technical specifications

7.1

Measuring interface

The 856 Conductivity Module has one measuring interface with one mea-
suring input.

The measuring cycle is 100 ms for all measuring modes.

7.1.1

Conductivity

One measuring input (Cond. Cell) for a 5-ring conductivity measuring cell
with integrated temperature sensor.

Measuring range

0 - 500 mS

Resolution

4 significant places

Measuring accu-
racy

Range 1 - 5 µS: ±1%; ±1 digit
Range 5 - 50 µS: ±0.5%; ±1 digit
Range 50 - 500 µS: ±0.5%; ±1 digit
Range 0.5 - 5 mS: ±0.5%; ±1 digit
Range 5 - 50 mS: ±0.5%; ±1 digit
Range 50 - 500 mS: ±1%; ±1 digit

7.1.2

Temperature

One measuring input (Cond. Cell) for a 5-ring conductivity measuring cell
with integrated Pt1000 temperature sensor.

Measuring range

–40 - +150 °C

Resolution

0.1 °C

Measuring accu-
racy

±0.2 °C
(typically ±0.1 °C; ±1 digit; without sensor error, under reference con-
ditions)