When the test is completed, the pop-up window
to the left appears. If the user would like a
printed record of the report, click Print Report,
and then Close Report.

16.4

Usp Wavelength Accuracy

The USP Wavelength Accuracy test measures wavelength uncertainty at three peaks in SRM-1920:
1261, 1681, and 1935nm.

From USP Test choose, Wavelength Accuracy.

The software will prompt for the type of
standard that will be used. If using the XDS
instrument, choose SRM-1920 with talc. For
Model 5000 and 6500 instruments choose
SRM-1920.

After choosing the type of standard that will be
used, the user must choose the standards file for
the wavelength standard that will be used. This
is on the diskette or CD-ROM that is supplied
with the standards set.