3.1 Equipping the eluent bottle

■■■■■■■■■■■■■■■■■■■■■■

6

■■■■■■■■

IC equipment: Additional eluent for 941

1 Mounting the adsorber tube

Fill the adsorber tube with adsorber material and cotton and insert it
into the M6 opening (2-1).

2 Connecting the FEP tubing

Fasten the tubing connection to the Dosino to the M6 opening (2-2).

Insert the FEP aspiration tubing into the M6 opening.

Screw the FEP tubing into the M6 opening.

3 Connecting the eluent aspiration tubing

Fasten the capillary connection to the ion chromatograph to the M8
opening (2-3).

Start by pushing the M8 tubing nipple over the end of the aspira-
tion tubing, followed by the O-ring.

Guide the eluent aspiration tubing through the M8 opening of
the bottle cap and temporarily secure it in place with the tubing
nipple.

Attach the tubing weighting and the aspiration filter to the end of
the eluent aspiration tubing, see the manual of the ion chro-
matograph
.

4 Attaching the bottle cap

Screw the bottle cap onto the eluent bottle.

Loosen the tubing nipple a little.

Push the eluent aspiration tubing so far into the eluent bottle that
the aspiration filter is at the bottom of the bottle.

Tighten the tubing nipple again.

5 Connecting the level sensor cable

Connect the two plugs of the level sensor cable to the two sockets
(2-4) and (2-5).