Metrohm Additional eluent for 941 User Manual

Page 17

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

IC equipment: Additional eluent for 941

■■■■■■■■

11

3 Connecting the Dosino to the ion chromatograph

NOTE

The 941 Eluent Production Module must be switched off when
the Dosino is being plugged to the MSB connector.

Check whether the 941 Eluent Production Module is switched on.
If so, switch off the instrument.

Plug the Dosino cable into one of the 941 Eluent Production
Module's MSB sockets.

Installing the FEP tubings on the Dosino

Accessories

For this step you need the following accessories:

800 Dosino (2.800.0010) with 807 Dosing Unit 50 mL (6.3032.250)
attached

The three FEP tubings connected to the bottles (6.1805.530),
one additional FEP tubing (6.1805.530)

Adapter Dosino Port 4 (6.1808.280)

Adapter UNF / M6 (6.2744.080)

1 Connecting the lateral ports

1

Concentrate

2

Port 1

3

Port 3

4

Eluent

Screw the concentration bottle's FEP tubing onto Port 1 of the
Dosino.

Screw the eluent bottle's FEP tubing onto Port 3 of the Dosino.