Metrohm 901 Titrando User Manual

Page 33

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

901 Titrando

■■■■■■■■

25

3

4 Adjust the height of the KF titration cell by pressing the locking lever.

It should almost touch the surface of the stirrer. The position can
now be fixed by readjusting the clamping ring.

Once the height of the KF titration cell has been adjusted correctly,
the entire cell can be raised and swiveled as required by pressing the
locking lever.

Filling the adsorber tube

Before insertion, the 6.1403.040 adsorber tube must be filled with the
6.2811.000 molecular sieve. Proceed as follows:

6.1403.040

1

2

3

6.2811.000

1 Insert a small cotton plug into the bottom of the adsorber tube. Do

not pack the cotton too tightly.

2 Fill the molecular sieve up to the ¾ level.

3 Place a small cotton plug on the molecular sieve. Do not pack the

cotton too tightly.

4 Seal the adsorber tube with the appropriate cover.