3 Installation

901 Titrando

■■■■■■■■

27

6 Screw the 6.1805.200 PTFE M8 tubing of the solvent bottle onto the

dosing tip.

Inserting electrode, adsorber tube and septum stoppers

6.2730.020

6.2730.010

6.0338.100

6.1403.040

6.2104.020

1

2

3

4

5

6

Proceed as follows:

1 Introduce the 6.2730.020 septum stopper (with septum inserted)

into the front opening of the vessel lid.

2 Insert the O-rings of the electrode and of the adsorber tube into the

middle openings of the vessel lid.

3 Screw the two 6.2730.010 screw nipples into the openings with the

O-rings. Do not screw too tightly.

4 Introduce the 6.0338.100 double Pt electrode into the left-hand

opening and then tighten the screw nipple until it seals.

5 Screw the 6.2104.020 electrode cable tightly onto the electrode.