3 indicator electrode – Metrohm 901 Titrando User Manual

Page 47

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Karl Fischer titration

901 Titrando

■■■■■■■■

39

4.1.7.3

Indicator electrode

A new indicator electrode can take a certain warm-up time for forming
the surface. During this time unexpectedly long titration times and high
measurement results can occur. This phenomenon will, however, disap-
pear after a short time of use. In order to accelerate the setting of a new
indicator electrode, the 901 Titrando can be conditioned e.g. over night.

A contaminated indicator electrode can be carefully cleaned with an abra-
sive agent (6.2802.000 polishing set or toothpaste). After the cleaning,
rinse with ethanol.

The two Pt wires of the indicator electrode should run as parallel as possi-
ble to one another. Check the Pt wires before inserting the electrode.