3 Installation
940 Professional IC Vario ONE/SeS/PP (2.940.1500)
■■■■■■■■
35
3.13.2.1
Installing the eluent path
The eluent path is connected with the capillaries in and out.
1 Connect the capillary labeled with in with the output of the separa-
tion column using a short pressure screw (6.2744.070).
2 Use a long pressure screw (6.2744.090) to connect the capillary
labeled with out to the input of the MCS (see Chapter 3.15.2, page
44).
3.13.2.2
Connecting the regeneration solution
The regeneration solution is connected to the regenerant capillary.
Connecting the regeneration solution to the peristaltic pump
For this step you need the following accessories:
■
Accessory kit: Flex/Vario: ChS (6.5000.030)
■
Pump tubing (6.1826.420)
■
Tubing olive with filter and lock (6.2744.180)
■
Tubing olive (6.2744.034)
■
Tubing cartridge of the peristaltic pump
1 Prepare a tubing cartridge of the peristaltic pump for the regenera-
tion solution (see Chapter 3.14.1, page 37).
2 Fasten the capillary labeled regenerant to the outlet of the pump
tubing using a pressure screw (6.2744.070).
3 Fasten the PTFE capillary from the bottle with the regeneration solu-
tion to the inlet of the pump tubing.