14 peristaltic pump, 1 installing the peristaltic pump, Installing the peristaltic pump – Metrohm 940 Professional IC Vario ONE/SeS/PP User Manual

Page 47

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

940 Professional IC Vario ONE/SeS/PP (2.940.1500)

■■■■■■■■

37

2 Fasten the capillary labeled rinsing solution to the outlet of the pump

tubing using a pressure screw (6.2744.070).

3 Fasten the PTFE capillary from the bottle with the rinsing solution to

the inlet of the pump tubing.

3.14

Peristaltic pump

3.14.1

Installing the peristaltic pump

Installing the pump tubing
Pump tubing can differ in terms of material, diameter and thus flow rate.
Different pump tubing is used depending on the application.

Table 2

Pump tubing

Order num-
ber

Name

Material

Inner diame-
ter

Use

6.1826.310

Pump tubing LFL (orange/
green), 3 stoppers

PVC (Tygon

®

)

0.38 mm

Pump tubing for bromate deter-
mination using the triiodide
method.

6.1826.320

Pump tubing LFL (orange/
yellow), 3 stoppers

PVC (Tygon

®

)

0.48 mm

For acceptor solutions for inline
dialysis and for inline ultrafiltra-
tion.

6.1826.330

Pump tubing LFL (orange/
white), 3 stoppers

PVC (Tygon

®

)

0.64 mm

No special applications.

6.1826.340

Pump tubing LFL (black/
black), 3 stoppers

PVC (Tygon

®

)

0.76 mm

For sample solution in inline dial-
ysis.

6.1826.360

Pump tubing LFL (white/
white), 3 stoppers

PVC (Tygon

®

)

1.02 mm

For sample transfer.

6.1826.380

Pump tubing LFL (gray/
gray), 3 stoppers

PVC (Tygon

®

)

1.25 mm

For inline sample dilution.

6.1826.390

Pump tubing LFL (yellow/
yellow), 3 stoppers

PVC (Tygon

®

)

1.37 mm

For sample solution in inline
ultrafiltration.

6.1826.420

Pump tubing PharMed

®

(orange/yellow), 3 stoppers

Ismaprene

0.51 mm

For suppressor solutions.

6.1826.020

Pump tubing (blue/blue), 2
stoppers

PVC (Tygon

®

ST)

1.65 mm

Pump tubing for online IC instru-
ments and automation in voltam-
metry.

Advertising