6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

110

■■■■■■■■

850 Professional IC – Anion MSM-HC – MCS – Prep 1

6 Troubleshooting

6.1

Problems and their solutions

Problem

Cause

Remedy

Marked drop in
pressure

Leak in the system.

Check and seal connections (see Chapter 3.5,
page 24)
.

Marked rise in pres-
sure

6.2821.120 inline filter
blocked.

Replace 6.2821.130 filter (see Chapter 5.6,
page 93)
.

MSM-HC – blocked.

Regenerate the MSM-HC (see Chapter
5.11.3.1, page 98)
.

Note: 6.2821.180 pump tubing connection
with filter must be used .

Conductivity detector
blocked.

Shorten capillary ends by several mm (see
Chapter 5.14.1, Page 107)
.

Rinse detector against the normal flow
direction (see Chapter 5.14.1, Page 107).

Guard column – blocked.

Replace guard column (see Chapter 3.21,
page 72)
.

Separation column –
blocked.

Regenerate separation column (see Chap-
ter 5.15.4, page 108)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 75)
.

Note: Samples should always be microfiltered
(see Chapter 5.7, page 95).

Injection valve – valve
blocked.

Have the valve cleaned (by Metrohm service
technicians).

Drift of the baseline

Thermal equilibrium not yet
attained.

Condition (see Chapter 3.15, page 54) instru-
ment with the column thermostat switched
on .

Leak in the system.

Check and seal connections (see Chapter 3.5,
page 24)
.