2 intended use, Intended use – Metrohm 850 Professional IC Anion MSM-HC MCS Prep 1 User Manual

Page 13

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

850 Professional IC – Anion MSM-HC – MCS – Prep 1

■■■■■■■■

3

interactions with the column. The Metrohm analytical columns are equip-
ped with a chip on which their technical specifications and their history
(first use / setting up, operating hours, injections, ... ) are saved.

1.2

Intended use

The 850 Professional IC – Anion MSM-HC – MCS – Prep 1 is used
for ion chromatographic determination of anions with sequential sup-
pression
and inline sample preparation:

Sequential suppression consists of:

Chemical suppression by the High Capacity Metrohm Suppressor Mod-
ule (MSM-HC) (see Chapter 3.16, page 57) and subsequent

CO

2

suppression by the Metrohm CO

2

Suppressor (MCS) (see Chapter

3.18, page 65).

The use of this technology reduces background conductivity to a mini-
mum and improves the detection accuracy.

If required, the instrument can also be used for the determination of cati-
ons and anions without suppression.

The peristaltic pump and the cell holder in the lower part of the instru-
ment make it possible to use various IC equipments for inline sample
preparation
, e.g.:

6.5330.000 IC equipment for dialysis

6.5330.010 IC equipment for ultrafiltration

The present instrument is suitable for processing chemicals and flammable
samples. The usage of the 850 Professional IC – Anion MSM-HC – MCS –
Prep 1 therefore requires that the user has basic knowledge and experi-
ence in the handling of toxic and caustic substances. Knowledge with
respect to the application of the fire prevention measures prescribed for
laboratories is also mandatory.