5.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

80

■■■■■■■■

850 Professional IC– Cation

Problem

Cause

Remedy

Vacuum is not being
built

Eluent Degasser – Connec-
tor Vacuum on the rear of
the instrument not (tightly)
sealed.

Seal the connector Vacuum tightly with a
6.144.6040 threaded stopper.
or when using an 872 Extension Module:

Connect an FEP tubing between the IC
instrument and the Extension Module and
tighten both clamping screws until they
seal.