Note: Only qualified electricians should be allowed to perform the installation of the

instrument as it involves a line voltage that could endanger life.

CAPTIVE SCREW

HIGH VOLTAGE COVER

ALARMS

BULKHEAD

COMMUNICATION
BULKHEAD

POWER
BULKHEAD

POWER

TERMINAL

BLOCK

POWER CABLE
STRAIN RELEIF

ANCHOR

ALARM #2
TERMINAL

BLOCK

ALARM #1
TERMINAL

BLOCK

RS-485

TERMINAL

BLOCK

4-20mA
TERMINAL

BLOCK


Figure 5: Electrical Connections for the Instrument

4.4.1 Power
The instrument is equipped with 100-240 VAC, 47-63 Hz power supplies requiring
150VA; please verify that the line voltage falls within these specifications. It is
recommended that a circuit breaker be placed prior to the power connection to
allow for service. For safety it is recommended that the connection be less than 2
meters (six feet) from the instrument. While making connections, refer to Figure 5.

The CLX is intended for cord connection with a three wire non-locking grounded
power cord. A power cord can be purchased separately from the factory (Catalog
No. 20779S). The CLX is not supplied with a power cord. If the CLX is to be
used in the U.S. or Canada the power cord must be UL Listed & CSA Certified.
Please consult all local electrical codes for proper connection.