Specifications – HF scientific CLX Online Residual Chlorine Monitor (Old) User Manual

Page 7

Advertising
background image

CLX (9/07)

Page 1

REV 3.1

Specifications

Measurement Range

0 – 10.00 mg/L (PPM)

Accuracy

±5% or ±0.03 mg/L (PPM) for range of 0-6 mg/L(PPM)

Resolution

0.01 mg/L (PPM)

Cycle Time

Adjustable: 90 seconds to 10

minutes (defaults to 2.5 minutes)

Display

Multi-Line Liquid Crystal Backlit Display

Alarms

Two Programmable, 120-240VAC 2A Form C Relay

Analog Output

Powered 4-20 mA, 600

Ω drive, isolated

Communications Port

Bi-directional RS-485 with Modbus

Maximum Water Pressure

Integral pressure regulator rated 1380kPa (200 PSI.)

Flow Rate

200 ml/min. minimum flow

Operating Temperature

5°C – 40°C (41°F – 104°F)

Wetted Materials

Nylon, Borosilicate Glass, Silicon, Polypropylene, Stainless Steel, Brass

Sample Temperature Range 5°C – 40°C (41°F – 104°F)

Power Supply

100

– 240 VAC, 47 – 63 Hz, 150VA

Insulation Rating

Double Insulated, Pollution Degree 2, Overvoltage Category II

Environmental Conditions

Not recommended for outdoor use.

Altitude up to 2000 meters

Up to 95 % RH (non-condensing)

Enclosure Rating

Designed to meet IP 66 /NEMA 4X

Regulatory Compliance

And Certifications

CE Approved, ETL listed to UL 3111-1 &

ETL Certified to CSA 22.2 No. 1010-1-92

Shipping Weight

2.5 kg (5.5 lbs.)

Shipping Dimensions

406 mm X 406 mm X 229 mm (16”X 16” X 9”)

Warranty

1 Year from date of shipment

Advertising