9.2 Clearing Faults

Every time SERVICE mode is exited, all faults are cleared. If the original fault or a new

fault occurs, it will be posted.

9.3 Diagnostic Chart

Symptom

Cause Cure

Lower display shows MA

4-20 mA loop open

Check wiring. See sections
4.4.4 and 7.2

Lower display shows FAIL

Major system fault

Refer to section 9.1 & 9.2

Readings are erratic

(1) Bubbles in solution
(2) Debris in flow

(1) See above
(2) Install T strainer at inlet

Readings are lower than expected (1) Condensate or leaky

measurement cuvette

(2) Measurement cuvette

dirty

(3) Reagents bad or expired

(1) Install desiccant cartridge

kit

(2) Replace or clean cuvette

(3) Replace reagents

Upper display flashes

Sample Over-Range

Check sample. Sample may be
too high to read.

Upper display shows nOnE while
attempting to calibrate

No current reading displayed

Wait for CLX to post a reading


9.4 Technical and Customer Assistance