0 instrument calibration, 1 slope (gain) calibration procedure – HF scientific CLX Online Residual Chlorine Monitor User Manual

Page 18

Advertising
background image

CLX (1/12)

Page 13

REV 5.1

6.0

Instrument Calibration

The instrument was tested prior to leaving the factory. Since it operates from a pre-
determined calibration curve, no calibration is required.

Calibration can easily be performed if required. The method is by comparison against
another instrument, such as a laboratory or hand held photometer (such as HF scientific’s
Chlorine Pocket Photometer).

There are two points of calibration. The slope or gain and the zero (offset). To perform the
zero the instrument must be plumbed to a sample of known chlorine free water such as de-
ionized water for a zero adjustment.

6.1 Slope (gain) Calibration Procedure
It is important that the chlorine level be quite stable to use this method. The comparison
will be made against a trusted measurement such as a chlorine photometer,
spectrophotometer, or an amperometric titration.

1. Obtain a grab sample of the flow.
2. Measure the value of the sample with one of the methods shown above.
3. On the CLX, press the MODE/EXIT button once. The screen is shown below.

4. Press  to enter the calibration adjustment.

5. The screen will show the current reading on the CLX. Using the  &  buttons

adjust the reading to agree with the laboratory method or portable photometer.

6. Press  to accept the calibration adjustment and return to AUTO measurement

mode.

There is a limit to the size of the change that can be made to a current
reading. The upper limit is the current reading times 1.5. The lower limit is
the current reading divided by 1.5.

Ensure a reading is posted to the display before calibrating to avoid a nOnE
error.

Advertising