Measurement range, Resolution – HF scientific CLX Online Residual Chlorine Monitor User Manual

Page 6

Advertising
background image

CLX (1/12)

Page 1

REV 5.1

Specifications

Measurement Range

0 – 10.00 mg/L (PPM)

Accuracy

±5% of reading or ±0.03 mg/L (PPM) whichever is greater for range of
0-6.0 mg/L(PPM)

±10% of reading from 6.01-10.00 mg/L (PPM)

Resolution

0.01 mg/L (PPM)

Cycle Time

Adjustable; 110 seconds to 10

minutes (600 seconds)

Note: the system defaults to 2.5 minutes

Display

Multi-Line Liquid Crystal Backlit Display

Alarms

Two Programmable, 120-240VAC 2A Form C Relay

Analog Output

Powered 4-20 mA, 600

Ω drive, isolated

Communications Port

Bi-directional RS-485 with Modbus

Water Pressure

Integral pressure regulator 0.34 bar(5.0 PSI) to 10.3 bar (150 PSI.)

Flow Rate to Waste

200 – 400 ml/min.

Operating Temperature

0°C – 40°C (32°F – 104°F)

Wetted Materials

PVC, Borosilicate Glass, Reslyn (FFKM), Viton

®

(FKM),

Polypropylene, Stainless Steel, Acetal, Noryl

®

, Silicone

Sample Temperature Range 0°C – 55°C (32°F – 131°F)

Operating temperature for 30 day reagent life is 0-40°C (32°F – 104°F)

Reagent life will be <30 days if kept at temperatures above 40°C

Power Supply

100

– 240 VAC, 47 – 63 Hz, 150VA

Insulation Rating

Double Insulated, Pollution Degree 2, Overvoltage Category II

Environmental Conditions

Not recommended for outdoor use.

Altitude up to 2000 meters

Up to 95 % RH (non-condensing)

Enclosure Rating

Designed to meet IP 66 /NEMA 4X

Regulatory Compliance

And Certifications

CE Approved, ETL listed to UL 61010-1-2004 &

ETL Certified to CSA 22.2 No. 61010.1 2

nd

edition dated July 2004

Shipping Weight

3.9 kg (8.6 lbs.) Reagents Shipped Separately

Shipping Dimensions

406 mm X 406 mm X 241 mm (16”X 16” X 9 ½”)

Warranty

2 Years from date of manufacture

Advertising