Sky-Watcher Synscan Hand Control User Manual

Page 18

Advertising
background image

34

35

PART XI : ADVANCED FUNCTIONS

The SynScan hand control divides the sky into 85 small zones, and users can calibrate the

pointing error for each of these zones. The next time that the SynScan controller tries to

locate an object in the calibrated zone (or a zone nearby), it will automatically apply the re-

corded calibration data to compensate the pointing error.
This function is useful for locating faint deep sky objects, and it is also helpful for obtaining
consistent pointing accuracy for a permanent observatory.

Here are the instructions on using the PAE function:

1.

Perform a 1-star alignment, 2-star alignment, or a 3-star alignment.

2.

Choose a celestial object in a zone of interest as reference by referring to a sky map or

planetarium software. In general, it is a rather bright star, but users can also use other

objects. Use the SynScan hand control to control the mount to point the telescope to the

reference object.

3.

Use one of the following operations to start the PAE calibration:

• Press the “

UTILITY

” shortcut key, access to sub-menu “

PAE\PAE Correction

”, and then

press the

ENTER

key.

• Press and hold the

ESC

key for two seconds.

4.

The screen will display “

Re-centering Obj.:

” in the first row, and display the name of the

reference object in the bottom row. (If the last object is launched from a PC, then instead

of the name of the reference object, it will display “

The last target

”.) Now use the direction

keys of the hand control to center the object in the telescope’s FOV, and then press

EN-

TER

to confirm. Remember to end the centering operation by pressing the

Right

and

Up

direction keys together.

5.

Repeat

Step 2

to

5

for viewing different portions of the sky.

Note:
• Whenever the SynScan hand control locates an object, it will automatically check whether

a PAE calibration data is available, and apply the compensation accordingly. No manual

intervention is required.

• If multiply PAE calibration is performed in the same zone, the previous calibration data

will be overwritten.

• Users can access the menu “

UTILITY FUNCTION > PAE > Clear PAE data

” to clear all PAE

calibration data.

• The PAE calibration data will be automatically cleared after a

1-star alignment, 2-star

alignment or 3-star alignment.

PART XI : ADVANCED FUNCTIONS

The polar alignment function can help users to polar align an equatorial mount accurately.

Here are the operating instructions:

11.3 Polar Alignment without Polar Scope

1.

Complete a 2-star alignment or a 3-star alignment. At the end of the alignment, the Syn-

Scan hand control will display the polar alignment error

(refer to Section 3.3)

. Users can

use the data to determine whether it is necessary to adjust the polar alignment.

2.

Press the “

MENU

” shortcut key, and then access to sub-menu “

Alignment\Polar Alignment

”, press the

ENTER

key to proceed to the next step.

3.

The screen will display “

Select a Star

”.

• Use the scroll keys to browse through a list of star names and press the

ENTER

key to

pick one as the reference star for polar alignment.

• The mount will start slewing to point the telescope to the reference star.

4.

Use the direction keys to center the reference star in the eyepiece of the telescope after

the mount stops slewing. Remember to end the centering operation with

Up

and

Right

direction keys. Press the

ENTER

key to proceed to the next step.

5.

The screen will now display the polar alignment error in altitude (Mel=dd°mm’ss”). Users

can then use the data to determine whether or not to adjust the altitude of the R.A. axis in

the next step. Press the ENTER key to proceed.

6.

The mount will slew to a new position. When it stops, the screen will display “Adjust

Altitude:”. By using ONLY the altitude control of the mount (do not touch the azimuth

control), bring the reference star back to the closest point to the center of the FOV of the

telescope’s eyepiece. Remember the reference star’s current position in the eyepiece for

later adjustment. Press the ENTER key to confirm the centering operation.

7.

The screen will now display the polar alignment error. Users can then use the data to

determine whether or not to adjust the azimuth of the mount in the next step. Press the

ENTER key again to proceed to the next step.

8.

The mount will slew to a new position. When it stops, the screen will display “Adjust Azi-

muth:” By using ONLY the azimuth control of the mount (do not touch the altitude control),

bring the reference star back to the closest point to the previous position (at the end of

Step 6). Press the ENTER key to confirm the centering operation.

9.

The screen will display the polar alignment error again, press the ENTER button to end

the polar alignment process.

10.

Go back to the “

Alignment

” menu on the SynScan hand control and execute another

2-Star or 3-Star alignment, and then check the polar alignment error data reported at the

end of the 2-star alignment or 3-star alignment. Repeat Step 2 to Step 9 until the error

is small enough and acceptable. Generally, users can get up to 1 arc-minute polar align-

ment accuracy after repeating this polar alignment process 2 or 3 times.

Note:
• Users can press the

ESC

key at any time during the polar alignment process to exit.